P821: OAKTON® pH and CONDUCTIVITY METERS IN PRINTING
Accurate mixing and monitoring of press fountain solutions are an integral part of quality printing. Changes in technology and raw materials present a constant challenge to today’s press operators. Oakton’s advanced design in portable, pocket-sized meters makes them a leader in the field of quality control in the lithographic industry.
The micro-processor based instrumentation of the pH meter makes it exceptionally fast, stable, accurate and repeatable. Eliminate the guessing associated with common, paper pH test strips. The economically priced pH Testr series allows them to be used at every press.
Maximum quality control necessitates frequent conductivity readings of the fountain solution. Contamination from paper coatings, ink and spray powder cannot be determined with pH readings. Only conductivity monitoring can pin-point problems before they occur.
How to use Oakton Testrs
- Read all instructions relating to the meters before attempting to use them.
- Once familiar with the meters, take a pH reading and a conductivity reading of the water that will be used to mix the fountain solution. Record these readings in a log book for reference.
- Measure one gallon of water in to a clean container.
- Add one ounce of fountain solution concentrate to the water.
- Take a second conductivity reading of the water.
- Subtract the initial conductivity reading from the second. The difference is the amount of conductivity that is added per ounce of the specific fountain concentrate being used. It will vary from solution to solution. Make note of this value.
- Divide 1000 by the value of conductivity added per ounce. The resulting answer represents the starting point for the number of ounces of fountain concentrate to use per gallon of water.
- Add the required fountain concentrate. The conductivity of this mixture should be about 1000 mmhos higher than the starting water. This condition is commonly referred to as “one thousand over water”. Remember, this is not an absolute number and it can vary as the water varies, or as the fountain concentrate changes.
- Record this conductivity. When monitoring the fountain solution, readings that are 500 mmhos or more than the starting point are a warning that problems may occur. Readings of 1000 mmhos or higher than the original are almost always associated with problems, and the fountain should be dumped and re-mixed.
- Finally, take a pH reading and record it for reference. During the press run, changes in the pH almost never occur more than ± 0.2 to 0.4 pH units. The actual pH value should be approximately between 3.8 to 5.5.
USING OAKTON METERS TO TROUBLESHOOT FOUNTAIN PROBLEMS
Problem | Cause | Solution |
---|---|---|
1. Toning, Tinting, Scumming An irregular film of ink printing in the non-image areas of a plate. | 1. Incorrect pH or conductivity. 2. Insufficient gum or wetting agents in fountain solution. 3. Plate improperly exposed or under-developed. | 1.Check and/or correct pH and conductivity as explained above. 2.Check mixture of fountain solution. 3.Use gray-scale when exposing plates; fully develop per directions. |
2. Ink emulsification Too much water in the ink. | 1. Incorrect pH or conductivity. 2. Fountain solution mixed too strongly. 3. Too much wetting agent in fountain solution. 4. Ink tack too low. | 1.Check and/or correct pH and conductivity as explained above. 2. Mix fountain solution as explained above. 3. Reduce gum, alcohol or alcohol replacement. 4. Add binding varnish to ink too thin from can or lower amount of reducing compound used on stiff ink. |
3. Roller stripping Ink rollers do not accept ink, usually in bands around the rollers. | 1. Fountain solution too acidic. 2. Running too much water. 3. Conductivity of starting water too high. 4. Glazed rollers. 5. Worn rollers. | 1. Check pH of fountain solution. Adjust as above. 2. Establish a new ink-water balance. 3. Use distilled, deionized or RO-purified water. 4. Deglaze rollers with hot water and deglazer. 5. Replace worn rollers. |
4. Ink not drying Ink remains wet too long, causing problems with offsetting, smearing, color trapping and perfecting. | 1. Fountain solution too acidic. 2. Running too much ink. 3. Ink not compatible with printing stock. 4. Ink has emulsified | 1. Balance fountain solution as explained above. 2. Reduce ink to almost zero, then build-up in small steps for new ink-water balance. 3. Match ink to stock. 4. See Ink Emulsification above. |
5. Plate blinding Plate is not accepting uniform ink film or not accepting ink at all. | 1. Fountain solution too acidic. 2. Too much gum in fountain solution. 3. Plate surface contaminated. 4. Too much water. | 1. Balance fountain solution with meters as explained above. 2. Mix fountain solution correctly. 3. Clean plate with quality plate cleaner, then rinse well with water. 4. Reduce water; establish new ink-water balance. |
6. Plate wear Image on plate wears or breaks-down prematurely. | 1. Fountain solution pH incorrect. 2. Too much pressure on plate from blanket or rollers. 3. Ink tack too high. 4. Running too much ink. | 1. Check and/or correct pH and conductivity as explained above. 2. Check and/or adjust packing and roller pressure settings. 3. Add reducing compound to ink. 4. Reduce ink settings and establish new ink-water balance. |
Manufacturer & master distributor
Lithco Products
From the Blog
PlateStrips™
Offers users of disposable printing plates the luxury of running tight register multi-color work without the added cost of film, stripping and metal plates. PlateStrips™ turns your disposable plates into metal…and gold for you! PlateStrips™ solve a real problem. Most printers are familiar with the problems associated with disposable printing plates. Paper and polyester plates…
Myron L Company pH and Conductivity Meters
The Myron L Company has become one of the world’s leading manufacturers of water quality control instruments. Their product philosophy of Accuracy, Reliability and Simplicity has established the Myron L meters as one of the standards of the printing industry. Combination pH/Conductivity meters make fountain solution monitoring a quick, one-step procedure.
Varn International Pressroom Chemistry
Varn Products is the number one name in pressroom chemistry today. Featuring ecologically formulated washes and fountain chemistry, Varn chemicals will enable you to meet new EPA regulations without sacrificing quality or drying speed.
Nacco Photo and Plate Chemistry
Nacco Chemicals manufactures premium quality photographic and plate developing chemicals at an economical price. Whether you are tray developing, using an imagesetter or running Silvermaster™ plates, Nacco can save you money while improving your quality.
Ihara Densitometers
Utilize advanced microcomputer technology to ensure superior performance and reliability for its entire line of densitometric products. Included within the line are B&W Reflection Densitometers, B&W Transmission Densitometers, Color Reflection Densitometers, Plate Densitometers, Combination B&W Reflection/Transmission Densitometers and Spectrophotometers. The microcomputer technology allows all reflection densitometers to feature quick calibration, menu driven commands and multi-lingual,…
LithoProof™ Blue Proofing Paper
Proof-It is currently available in Semi-Matte, Glossy, Matte, Heavyweight Matte, Publication Grade, Commercial Grade and Image Transfer for all models of Epson 3000 and 5000, Hewlet-Packard Design Jet and Encad Novajet. Make your Bluelines for less! Introducing LithoProof instant image blue proofing material. Now there is an economical and quality alternative to duPont Dylux®. LithoProof…